Global distribution and OEM cooperationsale@atombit.cn
BA121S
Water Sensor Interface ASICs

BA121S

Low-range Conductivity Interface ASIC

High-resolution low-conductivity measurement ASIC for purified-water and precision applications.

Downloads
Key Specifications
Interface
UART command control
Supply
3.3 V
Measurement Range
0–100 µS/cm

Product capabilities

Evaluate the capabilities that matter in your system

A low-range conductivity interface ASIC optimized for purified water and precision low-conductivity applications. The measurement core, NTC correction and UART output are integrated in SOP8.

01

0.1 µS/cm resolution

0–100 µS/cm range with 0.1 µS/cm resolution and error below 2% F.S.

02

Automatic temperature correction

0–100 °C temperature range with ±0.5 °C reference accuracy.

03

Single-channel conductivity measurement

3.3 V supply; ripple below 10 mV is specified for low-level measurement stability.

04

Bipolar drive reduces polarization

UART 9600 bps command interface and current below 3 mA.

Overview

Product overview

A low-range conductivity interface ASIC optimized for purified water and precision low-conductivity applications. The measurement core, NTC correction and UART output are integrated in SOP8.

High-resolution low-conductivity measurement ASIC for purified-water and precision applications. • 0.1 µS/cm resolution • Automatic temperature correction • Single-channel conductivity measurement • Bipolar drive reduces polarization • UART-controlled compact design

Engineering notes

Selection and system integration

  1. 010–100 µS/cm range with 0.1 µS/cm resolution and error below 2% F.S.
  2. 020–100 °C temperature range with ±0.5 °C reference accuracy.
  3. 033.3 V supply; ripple below 10 mV is specified for low-level measurement stability.
  4. 04UART 9600 bps command interface and current below 3 mA.

Technical data

Key Specifications

Use this information for initial selection; confirm final design and purchasing against the current approved datasheet and sample.

Interface
UART command control
Supply
3.3 V
Measurement Range
0–100 µS/cm
Accuracy
< 2% F.S.
Temperature
0–100 °C, ±0.5 °C

Application readiness

Installation, calibration and validation

Validate the product with real samples, final mechanics and the intended electrical environment before production.

  1. 01

    Treat power-supply noise, board leakage and connector contamination as part of the error budget.

  2. 02

    Use a probe and cell constant intended for low-conductivity water.

  3. 03

    Validate zero behavior and span using freshly prepared low-conductivity standards.

Public documents

Datasheets and product documents

Documents are based on AtomBit public product information; confirm the current revision before design release.

Start a project

Find the right sensing product for your project

Share the application, target parameters, installation space, interface and estimated quantity. Our team will recommend a product or customization route.

AtomBit Global Sales